IEC 61000-4-29
IEC 61000-4-29 DC Input Immunity Testing
Revision Edition 2.0, 2026-07 (published 30 July 2026), replaces Edition 1.0:2000
A single programmable DC power supply is an accepted generator architecture (Annex A, Figure A.2) — provided it holds 0.5–50 µs voltage-change edges into 100 Ω, sinks as well as sources peak inrush current to 220 A, keeps at least 20 % steady-state current margin over the EUT, and can present a genuine high-impedance output for interruptions: disconnecting, not merely programming 0 V, so the EUT cannot back-feed the gap.
This standard turns on
- voltage dips and interruptions on DC input ports
- high- versus low-impedance interruption modes
- durations in absolute seconds, not mains cycles — the DC counterpart of IEC 61000-4-11
And says nothing about
- ripple on the DC port — IEC 61000-4-17's job
- AC input ports — IEC 61000-4-11's job
- transient overvoltages injected by network faults — explicitly out of scope
- pass/fail severity selection — deliberately left to product committees
Cited limits
Every value carries the clause, table or figure it comes from.
| Parameter | Value | Clause |
|---|---|---|
| Voltage dip levels | 40 % and 70 % of UT (60 % and 30 % dips), abrupt microsecond-range transitions | Clause 5 / Table 1 |
| Short interruptions | 0 % UT for 0.001 / 0.003 / 0.01 / 0.03 / 0.1 / 0.3 / 1 s, in high-impedance and/or low-impedance condition | Table 2 |
| Voltage variations | 85 % and 120 %, or 80 % and 120 % of UT, for 0.1 / 0.3 / 1 / 3 / 10 s; ramp times set by the product standard | Table 3 |
| Rated-voltage-range rule | if the upper limit of the EUT's rated range exceeds the lower by more than 20 %, the full procedure runs at both limits | Clause 5 |
| Generator output voltage | up to 120 % of the maximum UT; a 780 V DC / 25 A generator is recommended to cover a large number of applications | §6.2.1 |
| Voltage-change edges | rise and fall 0.5–50 µs into a 100 Ω load, overshoot/undershoot < 10 % of the change | §6.2.1 |
| Peak inrush current drive | 50 A at 24 V, 100 A at 48 V, 220 A at 110 V, 100 A at 650 V — positive towards the EUT and negative reverse from it | §6.2.2 |
| Steady-state current margin | generator rated output current at least 120 % of the EUT's rated input current | §6.2.1 |
Where these figures came from
Each record below backs a specific part of this page. Where a table sits behind a paywall we say so above rather than reproducing it.
- IEC 61000-4-29:2026 — catalogue listing and abstract — iTeh Standards (IEC preview distributor). Retrieved 2026-08-24. Supports: edition, scope, list of Edition 2.0 technical changes, 61000-4-17 ripple boundary, §6.2.2 inrush values including 100 A at 650 V
- IEC 61000-4-29:2026 official preview sample (Edition 2.0 2026-07) — IEC, via iTeh Standards. Retrieved 2026-08-24. Supports: edition and date, contents, foreword, scope, supersession of the 2000 first edition
- REDLINE IEC 61000-4-29:2026 RLV preview sample — IEC, via iTeh Standards. Retrieved 2026-08-24. Supports: Clause 5 test levels, Table 2 and Table 3 durations, Clause 6 generator specifications, Annex A generator architectures, high/low impedance definitions
- IEC 61000-4-29 Edition 2.0 2026-07 public preview — IEC, via en-standard.eu. Retrieved 2026-08-24. Supports: scope, bibliography confirming the IEC 61000-4-11 and 61000-4-17 boundaries, publication date 30 July 2026 (also NSAI shop listing)