IEC 61000-4-11 / -4-13 / -4-14

IEC 61000 AC Immunity Testing

AC source with 500 A peak inrush drive, 1–5 µs dip edges, phase-of-dip control, and full per-phase arbitrary waveform synthesis with independent harmonic and interharmonic amplitude and phase control.

This standard turns on

  • voltage dips and interruptions
  • inrush drive capability
  • harmonic and interharmonic synthesis

And says nothing about

  • 400 Hz
  • DC bus transients
  • reliability duration testing

Cited limits

Every value carries the clause, table or figure it comes from.

ParameterValueClause
Dips, Class 20%/½ cycle, 0%/1 cycle, 70%/25 (50 Hz) or 30 (60 Hz) cycles61000-4-11 Table 1
Dips, Class 3adds 40%/10–12 cycles and 80%/250–300 cycles61000-4-11 Table 1
Interruptions0% for 250/300 cycles61000-4-11 Table 2
Generator rise/fall1–5 µs61000-4-11 §6
Peak inrush drive500 A (220–240 V) / 250 A (100–120 V)61000-4-11 §6
Harmonic rangeto 2 kHz (50 Hz mains) / 2.4 kHz (60 Hz mains)61000-4-13
Voltage fluctuationClass 2 ΔU ±8% U_n; Class 3 ΔU ±12% U_n61000-4-14
Scope — equipment currentUp to 16 A per phase; above that, IEC 61000-4-34 applies61000-4-11 §1
Current capability23 A at 70% U_nom, 40 A at 40% U_nom61000-4-11 §6