IEC 61000-4-11 / -4-13 / -4-14
IEC 61000 AC Immunity Testing
AC source with 500 A peak inrush drive, 1–5 µs dip edges, phase-of-dip control, and full per-phase arbitrary waveform synthesis with independent harmonic and interharmonic amplitude and phase control.
This standard turns on
- voltage dips and interruptions
- inrush drive capability
- harmonic and interharmonic synthesis
And says nothing about
- 400 Hz
- DC bus transients
- reliability duration testing
Cited limits
Every value carries the clause, table or figure it comes from.
| Parameter | Value | Clause |
|---|---|---|
| Dips, Class 2 | 0%/½ cycle, 0%/1 cycle, 70%/25 (50 Hz) or 30 (60 Hz) cycles | 61000-4-11 Table 1 |
| Dips, Class 3 | adds 40%/10–12 cycles and 80%/250–300 cycles | 61000-4-11 Table 1 |
| Interruptions | 0% for 250/300 cycles | 61000-4-11 Table 2 |
| Generator rise/fall | 1–5 µs | 61000-4-11 §6 |
| Peak inrush drive | 500 A (220–240 V) / 250 A (100–120 V) | 61000-4-11 §6 |
| Harmonic range | to 2 kHz (50 Hz mains) / 2.4 kHz (60 Hz mains) | 61000-4-13 |
| Voltage fluctuation | Class 2 ΔU ±8% U_n; Class 3 ΔU ±12% U_n | 61000-4-14 |
| Scope — equipment current | Up to 16 A per phase; above that, IEC 61000-4-34 applies | 61000-4-11 §1 |
| Current capability | 23 A at 70% U_nom, 40 A at 40% U_nom | 61000-4-11 §6 |
Further reading
- IEC 60601-1-2 Edition 4.1: Voltage Dips Are Now a Two-Voltage Test3 min · 29 cited facts
- IEC 61000-4-13: Synthesising Distortion on Command — and the Tables We Do Not Reprint4 min · 20 cited facts
- IEC 61000-4-29 Edition 2.0: One DC Supply as the Whole Generator4 min · 39 cited facts
- Assembling a Medical Power-Input Bench: The Generator Is the Specification4 min · 34 cited facts
- IEC 61000-4-11: The Inrush Requirement That Disqualifies Most AC Sources3 min · 21 cited facts